Multi-parameter Patient Monitor UP-8000 User Manual
User Manual for Multi-parameter Patient Monitor 3 1.6 Impact on the Environment and Resources Low 1.7 Safety Conform to IEC 60601-1, classified as Cla
User Manual for Multi-parameter Patient Monitor 4 Chapter 2 Working Theories of the Main Unit 2.1 Architecture The architecture of this monitor is s
User Manual for Multi-parameter Patient Monitor 5 4. The SpO2 module collects the data of pulse rate, pulse oxygen saturation (SpO2) and SpO2 volum
User Manual for Multi-parameter Patient Monitor 6 Chapter 3 Installation and Connection 3.1 Appearance 3.1.1 Front Panel Figure 3.1 Front Panel 1.
User Manual for Multi-parameter Patient Monitor 7 Press it to start or stop NIBP measure. 8. Print Press it to print different waveforms of differen
User Manual for Multi-parameter Patient Monitor 8 With Type CF applied part and applicable during the defibrillator is used
User Manual for Multi-parameter Patient Monitor 9 3.2.1 Description of Keys 3.2.2 Introduction for Display Screen The CSM always starts with display
User Manual for Multi-parameter Patient Monitor 10 Fig. B 3. Display Mode C Sensor impedance, CSI, BS% and operation time. When display C is active,
User Manual for Multi-parameter Patient Monitor 11 How to select menu settings from Display Mode 1. Remove battery lid. 2. From any Display Mode,
User Manual for Multi-parameter Patient Monitor 12 Info: Serial number Firmware version CSI version AD version Link: - indic
User Manual for Multi-parameter Patient Monitor I This Manual is written and compiled in accordance with the IEC 60601-1(Medical electrical equipment
User Manual for Multi-parameter Patient Monitor 13 3.3 Installation 3.3.1 Opening the Box and Check 1. Open the packaging, take out monitor accessorie
User Manual for Multi-parameter Patient Monitor 14 signals. 1. Connect the cable to the right-panel connector marked with the ECG icon. 2. Select elec
User Manual for Multi-parameter Patient Monitor 15 Transient caused by cable circuitry blocks while monitoring may be similar to the real heartbeat
User Manual for Multi-parameter Patient Monitor 16 Safety Instructions for NIBP Monitoring When taking the measure of an infant or neonate’s (les
User Manual for Multi-parameter Patient Monitor 17 install the SpO2 sensor on the finger with edema or fragile tissue. Do NOT put the SPO2 sensor
User Manual for Multi-parameter Patient Monitor 18 The conductive parts of sensors and their connectors, including the neutral sensor, should not cont
User Manual for Multi-parameter Patient Monitor 19 Chapter 4 Monitoring Screen 4.1 Main Screen 4.1.1 Date and Time Setup Instead of entering into moni
User Manual for Multi-parameter Patient Monitor 20 4.1.2 Screen Description Figure 4.2 Main Screen Border area Figure 4.3 Status/Information Area
User Manual for Multi-parameter Patient Monitor 21 Waveform area 1st Waveform: The first waveform is ECG waveform for lead II. Any lead can be sele
User Manual for Multi-parameter Patient Monitor 22 Figure 4.6 SpO2 and Pulse Rate Data Area “SpO2”: SpO2 label,The “99%” is the current SpO2 value
User Manual for Multi-parameter Patient Monitor II Instructions to User Dear Users, Thank you very much for purchasing our product. Please read the fo
User Manual for Multi-parameter Patient Monitor 23 Figure 4.9 Temperature Data Area TEMP1/2: Temperature label. The value below“36.7, 37.2” are t
User Manual for Multi-parameter Patient Monitor 24 4.2.1 Screen Description Freezing, ST segment analysis screen is similar with the main screen, exce
User Manual for Multi-parameter Patient Monitor 25 Last, press the Navigation Knob again. The screen shows “ST+0.xxx mV, Set ST, Dirc Ver”. Rotate the
User Manual for Multi-parameter Patient Monitor 26 4.4.1 Screen Description In the SpO2 list title bar, the front color of SpO2, PR is the same with S
User Manual for Multi-parameter Patient Monitor 27 4.6 Graphic Trend Screen 4.6.1 Screen Description Move the gray cursor to “TREND” in the mode selec
User Manual for Multi-parameter Patient Monitor 28 towards the other direction, the interval becomes 1 of the other direction. Therefore, it is very e
User Manual for Multi-parameter Patient Monitor 29 Figure 4.19 NIBP Graphic Trend Figure 4.20 SpO2 Trend Graph Figure 4.21 Respiration Graphic Tren
User Manual for Multi-parameter Patient Monitor 30 Figure 4.22 CSI Graphic Trend 4.6.2 Operation Instruction Rotate the Navigation knob to select par
User Manual for Multi-parameter Patient Monitor 31 Recall: press the recall button, the first record of recall list turns to green, rotate the knob to
User Manual for Multi-parameter Patient Monitor 32 During waveform recall, the system not only displays the current recalled waveform, but also displa
User Manual for Multi-parameter Patient Monitor III and standards as the accessories along with the monitor which are provided by the same factory, ot
User Manual for Multi-parameter Patient Monitor 33 Start: This button is used to start and end the system arrhythmia detection. The default is OFF. Wh
User Manual for Multi-parameter Patient Monitor 34 The Default button is used to resume default vales of parameter, but the Document information and i
User Manual for Multi-parameter Patient Monitor 35 Printer: Printer switch. If the printer option is “ON”, the printer will be available. This param
User Manual for Multi-parameter Patient Monitor 36 DIA: Diagnosis. No filtering, represent the true ECG with out filtering. OPE: Operation. Deep fil
User Manual for Multi-parameter Patient Monitor 37 TD:For setting temperature alarm range, if the temperature difference is higher than the setting
User Manual for Multi-parameter Patient Monitor 38 4.9.6 SpO2 Parameter Settings SpO2 Hi/Lo: high and low limits of SpO2 alarm Pulse Hi/Lo: high and
User Manual for Multi-parameter Patient Monitor 39 4.9.8 Resume Default System parameters default key. Move the cursor to “Default”, press the Navigat
User Manual for Multi-parameter Patient Monitor 40 ID: or Patient ID. To enter patient ID, choose the patient ID field using the Navigation Knob. Pres
User Manual for Multi-parameter Patient Monitor 41 Chapter 5 Technical Specifications 5.1 Cerebral State Monitoring 1. EEG sensitivity: ±400μV 2. Nois
User Manual for Multi-parameter Patient Monitor 42 5.4 TEMP Monitoring 1. TEMP measuring range: 25.0 ℃~45.0 ℃ 2. TEMP measuring accuracy: ±0.2℃ 3.
User Manual for Multi-parameter Patient Monitor IV Table of Contents CHAPTER 1 OVERVIEW ...
User Manual for Multi-parameter Patient Monitor 43 4. Hysteresis: ≤0.5mm 5. Frequency response: Monitoring mode: 0.5~40Hz Diagnostic mode: 0.05~75
User Manual for Multi-parameter Patient Monitor 44 (20) Sandpaper One piece Note: The accessories are subject to change. Detailed items and quantit
User Manual for Multi-parameter Patient Monitor 45 Chapter 7 Parameters Monitoring 7.1 CSM Monitoring 7.1.1 The Principle for Measurement The monito
User Manual for Multi-parameter Patient Monitor 46 EMG Facialis EMG may increase the CSI under certain conditions. The monitor incorporates an EM
User Manual for Multi-parameter Patient Monitor 47 Sensor Alarm The Sensor Alarm signals the interruption of reliable CSI calculation due to corrup
User Manual for Multi-parameter Patient Monitor 48 First of all, the hospital should be equipped with a 100~250V power supply system with a typical g
User Manual for Multi-parameter Patient Monitor 49 the oscillating method, the blood pressure at the maximum amplitude of cuff pressure is defined as
User Manual for Multi-parameter Patient Monitor 50 4. The studies have shown that the Korotkoff Sound Method has the worst accuracy when it comes to
User Manual for Multi-parameter Patient Monitor 51 6. Subject with hyperadiposis; In addition, statistics show that 37% people report blood pressur
User Manual for Multi-parameter Patient Monitor 52 2. Make sure that capillary arterial vessel beneath the finger is penetrated through by red and
User Manual for Multi-parameter Patient Monitor V 4.4.1 Screen Description ...
User Manual for Multi-parameter Patient Monitor 53 Chapter 8 Troubleshooting Note: In case of trouble of this machine in the service, follow the instr
User Manual for Multi-parameter Patient Monitor 54 Communication POWER indicator on CSM Link does not turn on when power is connected: Check mains
User Manual for Multi-parameter Patient Monitor 55 Chapter 9 Maintenance Note: In case of trouble of this machine in the service, follow the instructi
User Manual for Multi-parameter Patient Monitor 56 In order to avoid damaging the battery, do NOT use other power supply device to charge the batte
User Manual for Multi-parameter Patient Monitor 57 9.4 Storage If the equipment will not be used for long period of time, wipe it clean and keep it in
User Manual for Multi-parameter Patient Monitor 58 Appendix Abbreviation of Arrhythmia 1. ECG TACHY 2. ECG BRADY 3. ECG ARREST 4. MISS BEAT 5.
User Manual for Multi-parameter Patient Monitor 59 Other Terms and Messages “Cuff error” —cuff is not wrapped correctly, or is not connected
User Manual for Multi-parameter Patient Monitor 60 Default Alarming Values of All Parameters and Setup Range The default alarming value:
User Manual for Multi-parameter Patient Monitor 61 The high and low limits setting range: Type Parameter Adult Infant Neonate Hea
User Manual for Multi-parameter Patient Monitor VI 7.4.2 Points to be Noted in SpO2 and Pulse Measuring ...
Chapter 1 Overview 1.1 Features UP-8000 Multi-parameter Patient Monitor is the combination of the functions of Cerebral Status Monitor and Multi-para
User Manual for Multi-parameter Patient Monitor 2 waveforms; Resistance against defibrillator and electrosurgical unit interference and high safe